SONNET 3
CI and EAS Audio Processor
These support guides will help you handle and care for the SONNET 3 & SONNET 3 EAS Audio Processor.
Indicator Light
Confirmation

Status
Remote control command has been successfully received.

Status
Standby mode successfully activated.

Status
Direct wireless link to second audio processor successfully activated.
Program Change

Status
Program 1 is selected.

Status
Program 2 is selected.

Status
Program 3 is selected.

Status
Program 4 is selected.
Status

Status
If the optional status light has been activated: SONNET 3 is initialized and working.
Error Pattern

Problem
Temporary error or electronic malfunction
Solution
1. Switch off SONNET 3.
2. Switch on SONNET 3.

Problem
The setting you selected has not yet been programed, or there has been a program malfunction.
Solution
Select another setting.

Problem
Temporary error or electronic malfunction.
Solution
1. Switch off SONNET 3.
2. Switch on SONNET 3.

Problem
Programming failure, temporary error, or electronic malfunction.
Solution
1. Switch off SONNET 3.
2. Switch on SONNET 3.
Warning

Status
Batteries are empty.
Solution
Change the batteries.

Status
Maximum or minimum value of volume or audio sensitivity range has been reached.
Solution
Adjust volume or sensitivity if necessary.

Status
SONNET 3 is ready to perform an extended implant check.
Solution
Position the coil over the implant.
Flight Mode Confirmation Pattern

Status
Flight mode successfully activated
Coil Indicator Light
Confirmation

Status
The coil, coil cable, and SONNET 3 are working; your SONNET 3 and implant are successfully connected (applicable to SYNCHRONY series, CONCERTO series, SONATA series, and PULSAR implants).

Status
The coil, coil cable, and SONNET 3 are working (applicable to C40+ and C40 implants).
No Light Signal

Problem
Your audio processor is not working.
Solution
1. Change the batteries. If that does not solve the problem, go to 2.
2. Replace the coil cable. If that does not solve the problem, contact your CI center for assistance.
Problem
The indicator light has been deactivated by your audiologist.
Solution
Contact your audiologist.
Warning

Problem
The coil and implant are disconnected.
Solution
Position the coil over the implant.
Problem
SONNET 3 is in microphone monitoring mode.
Solution
Switch SONNET 3 off and on again.
Problem
The coil is positioned over the wrong implant.
Solution
Position the DL-Coil over the correct implant.
Problem
Coil cable is broken.
Solution
Replace the coil cable.
Problem
Batteries are empty.
Solution
Change SONNET 3’s batteries.

Problem
The coil has powered off.
Solution
1. Switch your SONNET 3 off.
2. Switch your SONNET 3 back on.
3. Reposition the coil over the implant.

Random signals
Problem
The coil cable is defective.
Solution
Replace the coil cable.

Refer to your user manual for more information.